Configure device options and preferences
2.
5.2
Enable or disable the Acknowledge All Alerts display
command
Display
ProLink III
Field Communicator Configure > Manual Setup > Display > Display Menus > Acknowledge All
Overview
You can configure whether or not the operator can use a single command to acknowledge
all alerts from the display.
Procedure
1.
2.
78
Option
Description
Disabled (de-
The display shows Display Variable 1 and does not scroll automatically. The
fault)
operator can move to the next display variable at any time using Scroll.
If you enabled Auto Scroll, set Scroll Rate as desired.
The default value is 10 seconds.
Tip
Scroll Rate may not be available until you apply Auto Scroll.
OFF-LINE MAINT > OFF-LINE CONFG > DISPLAY > ACK
Device Tools > Configuration > Transmitter Display > Ack All
Ensure that the alert menu is accessible from the display.
To acknowledge alerts from the display, operators must have access to the alert
menu.
Enable or disable Acknowledge All Alerts as desired.
Option
Description
Enabled (default) Operators can use a single display command to acknowledge all alerts at
once.
Operators cannot acknowledge all alerts at once. Each alert must be ac-
Disabled
knowledged separately.
Micro Motion
®
Compact Density Meters (CDM)