Mid Flow Computer And Power Supply Requirements; Flow Computer Density Calculations - Emerson Micro Motion CDM100M A Series Configuration And Use Manual

Compact density meters
Table of Contents

Advertisement

MID applications
Figure F-1:
A.
As a result of this test and certification, the CDM can be used in an MID liquid
measurement system.
F.2
MID flow computer and power supply
requirements
Approved flow computers
When using the CDM in an MID application, the CDM must use an approved flow
computer, such as a FloBoss S600 (TC7379, TC7470, TC8218), or an OMNI 3000/6000
(TC7375).
An approved flow computer is a flow computer which has either:
Power supply requirements
When not powered by a flow computer, the CDM must be powered by one of the following
power-supply types:
F.3

Flow computer density calculations

If you want to calculate temperature-corrected and pressure-corrected densities, you will
need external temperature and external pressure inputs.
218
MID label
Certificate number
An EC-type examination certificate for annex MI-005
An evaluation certificate or parts certificate issued by a notified body under the
Welmec 8.8 guideline
DC/DC Phoenix QUINT PS/24VCD/24VDC/5
AC/DC Phoenix QUINT PS/1AC/24VDC/3.5