RFL100 User Guide
Error code
Cause
Err 110
If using one probe with the data logger:
• No probe connected at startup
• Incompatible probe detected
• Trying to switch from two probe use to
single probe use without performing a
factory reset
If using two temperature probes with the
probe splitter:
• No probes connected at startup
• Two probes connected at same time
without following proper setup or
probe swap procedure
• Data logger is in dual probe mode but
one probe is missing
• Unsupported probe combination
• At least one incompatible probe
detected
Err 200
Real-time clock battery voltage is low.
Err 202
Communication failure with at least one
probe that was detected at startup.
Err 203
Probe error. Can be caused by probe
damage or a wet humidity sensor.
Err 204
Real-time clock temperature
compensation problem.
More information
‣ Probe detection (page 31)
‣ Probe replacement (page 39)
‣ Changing RFL100 clock battery (page 54)
58
Recommended action
1. Verify that a compatible probe is
connected to the data logger.
2. Power cycle the data logger to re-
detect the probe.
3. If the error persists, it is likely that the
data logger has been set up in two
probe mode. Perform the factory
reset procedure to allow single probe
use again, then power cycle the
logger to re-detect the probe.
1. Verify that that probe splitter and
two supported temperature-only
probes are connected to the data
logger.
2. If you reconnected any probes, power
cycle the data logger to re-detect the
probe(s).
3. If the error persists, perform the two-
probe setup again.
Replace the clock battery.
1. Check that the probe(s) are
connected properly.
2. If you reconnected any probes, wait
one minute for the display to update
and verify that the error is gone.
3. If the error persists, power cycle the
data logger to re-detect the probe(s).
Inspect the probe and replace it if
necessary. If the error has been caused
by a wet humidity sensor, wait for it to
dry out.
If error code Err 200 is also active,
replace the clock battery. If the error
persists, contact Vaisala.