5. To replace a cabled probe:
1
a. Loosen the locking ring of the connector at the end of the probe cable, and pull the
old probe away from the connector.
b. Connect the new probe and tighten the locking ring.
6. Move the power switch to the On position.
7. Look at the display and wait for the data logger to start up.
Data logger shows the word NEW on each affected channel when it detects that a probe
has been changed
the measurement reading(s). Verify that the display shows measurement readings instead
of dashes or error codes.

8. Close the battery cover of the data logger. Push the latch down until you hear a click. If

the cover does not close easily, push the probe (or extension cable) in and try again.
9. Insert the data logger back in the mounting bracket.