Table 18 Hmp110/T Probe Measurement Performance - Vaisala VaiNet RFL100 User Manual

Wireless
Hide thumbs Also See for VaiNet RFL100:
Table of Contents

Advertisement

Property
With mounting bracket
Weight
With batteries (2 pcs alkaline) and HMP115 probe 190 g (6.7 oz)
With batteries (2 pcs alkaline), HMP115 probe,
and magnetic mounting bracket
Materials
Housing
Display window
Sealings
7.2 RFL100 compatible probes technical
specification
Table 18 HMP110/T probe measurement performance
Property
Relative humidity
Measurement range
Accuracy in temperature range 0 ... +40 °C (+32 ... +104 °F)
0 ... 90 %RH
90 ... 100 %RH
Accuracy in temperature range −40 ... 0 °C, +40 ... +80 °C (−40 ... +32 °F, +104 ... +176 °F)
0 ... 90 %RH
90 ... 100 %RH
Factory calibration uncertainty at +20 °C (68 °F)
0 ... 90 %RH
90 ... 100 %RH
Humidity sensor
Stability
Temperature
Measurement range
Accuracy over temperature range
Specification
186 × 68 × 36.5 mm (7.32 × 2.68 × 1.44 in)
254 g (8.96 oz)
PC/ABS blend
PMMA (acrylic)
TPE
Description/Value
0 ... 100 %RH
1)
±1.5 %RH
±2.5 %RH
±3.0 %RH
±4.0 %RH
2)
±1.1 %RH
±1.8 %RH
HUMICAPâ 180R
±2 %RH over 2 years
−40 ... +80 °C (−40 °F ... +176 °F)
Chapter 7 – Technical data
1)
63

Advertisement