RFL100 User Guide
5.6 Changing RFL100 clock battery
• New 3 V lithium battery (type CR1/3N button cell)
• Small flat-head screwdriver
1. Open the battery cover of the data logger.
2. Use a small flat-head screwdriver to lift the top part of the small plastic cover marked
Clock battery (CR1/3N), and slide the cover upward until it comes loose.
3. Use the small screwdriver to lift the old clock battery from the battery socket.
4. Take the new clock battery and verify the + and - markings of the battery itself and the
clock battery socket. Insert the new clock battery in the clock battery socket.
5. Replace the clock battery cover.
6. Close the battery cover of the data logger. Push the latch down until you hear a click. If
the cover does not close easily, push the probe (or extension cable) in and try again.
5.7 Updating RFL100 firmware
Required:
• Computer with a free USB port and an operating system that supports the
Media Transfer Protocol (MTP). For example, Windowsâ 7 and newer.

• USB connection cable (USB 2.0 type A - micro-B, Vaisala cable 244961). You

can also use a generic cable that has all pins connected (not just power).
• RFL100 firmware update file from Vaisala
CAUTION!
Before updating, verify from the viewLinc Enterprise Server that up-to-date data
from this device is available. You can also copy the data from the device; see
Downloading data using service port (page
Updating the firmware will not affect the data logger's status in the Vaisala viewLinc
Monitoring System. If the device was accepted in the system before the update, it will remain
accepted.
1. Before starting the firmware update, read the release notes that are included with the file.
The release notes contain a summary of important changes and possible compatibility
requirements.
2. Turn the RFL100 on. The currently installed firmware version is shown briefly during start-
up. If the firmware version you have downloaded is newer than the installed version,
continue with the update.
54
Updating RFL100 firmware erases the recorded data on the device.
59).