Table 12 Rfl100 Error Codes - Vaisala VaiNet RFL100 User Manual

Wireless
Hide thumbs Also See for VaiNet RFL100:
Table of Contents

Advertisement

Problem
RFL100 turns off by itself.
6.2 Error codes
Table 12 RFL100 error codes
Error code
Cause
Err 100
User parameter bank checksum failure.
Err 101
Factory parameter bank checksum
failure.
Err 102
Real-time clock of the data logger has
lost accurate time.
Err 103
Main battery voltage is critically low.
Data logger has stopped radio
communication to conserve energy but
continues to record measurement data in
the local memory.
Err 104
Incorrect factory configuration
parameters.
Err 105
Real-time clock hardware error.
Possible cause
Display and LED of the RFL100
have been turned off remotely
using viewLinc Enterprise
Server. All other functions of
the data logger remain active.
Main batteries are empty.
Chapter 6 – Troubleshooting
Solution
You can change the setting
from the Hosts and Devices
tree in viewLinc Enterprise
Server.
If you are connecting the
RFL100 to a new viewLinc
Monitoring System and you
want the display and LED back
on again, press the Release
button of the RFL100 and
verify that the message NWK
REL appears on screen.
Replace the main batteries. See
Changing RFL100 batteries
(page
53).
Recommended action
Power cycle the data logger. If the error
persists, contact Vaisala.
Restore the wireless connection to an
AP10 access point. RFL100 will
synchronize its clock with the time from
the access point.
Replace the main batteries of the data
logger.
Power cycle the data logger. If the error
persists, contact Vaisala.