RFL100 User Guide
2.6 Alarm indicators
Figure 6 Alarm indicators on RFL100 display
1
Alarm indicators for channel 1
2
Alarm indicators for channel 2
Table 7 Alarm symbols
Symbol on display
RFL100 can show active threshold alarms on its local display. When a threshold alarm is active
on RFL100, the appropriate alarm indicators will be shown on the display. Additionally, the LED
will flash red for high-high and low-low threshold alarms.
Threshold alarms cannot be configured locally on the RFL100 itself; they are configured using
viewLinc Enterprise Server software. When applying a threshold alarm template to a Location,
you can choose to show the alarms on the data logger that is linked to the location. To show
the alarms, enable the Send to device setting, and then enable Alarm on Device for each
threshold that you want to generate an alarm on the RFL100.
On the RFL100, only one set of thresholds can be active at a time for one channel. The latest
set that is pushed to the device replaces the previous one. The Send to device setting of any
previously sent threshold alarm is automatically set to No.
16
1
2
Description
High-high threshold alarm active.
High threshold alarm active.
Alarm bell symbol that is always shown when any threshold
alarm is active on this channel.
Low threshold alarm active.
Low-low threshold alarm active.
M211861EN-D