11. To perform the adjustment using two reference environments (two-point adjustment),
perform these steps:
a. Select 2-point adjustment > SELECT > OK.
b. When the measurement is stable, select READY.
c. Give the reference RH value by using the arrow buttons and select OK.
d. Insert the probe to be calibrated in the second reference environment.
e. When the measurement is stable, select READY.
f. Give the reference RH value by using the arrow buttons and select OK.
g. To confirm the adjustment, select YES. If you select NO, you return to the adjustment
mode display and no changes are made.
h. Continue from
12. To perform the adjustment using a reference probe, perform these steps:
a. Select To same as RH[II].
b. To confirm the adjustment, select YES. If you select NO, you return to the adjustment
mode display and no changes are made.
13. Calibration and adjustment is now completed. Select BACK to exit the adjustment mode
and EXIT to return to the basic display.
14. Disconnect the calibrated probe from the MI70 indicator.
15. Reconnect the probe to the data logger.
More information
‣ Probe replacement (page 39)
step