12. Select [5] Note to edit the calibration info text that will be stored in the probe when the
adjustment is applied in
When done, select the OK character in the bottom right corner. To exit without changing
the text, select Cancel.
13. Select [6] Apply to view the calibration result. Verify the corrections in the confirmation
screen and select Apply to apply the adjustment to the probe, or Cancel to exit without
applying the adjustment.
14. Disconnect the probe from the HM40.
15. Connect the probe to the data logger.
More information
‣ Probe replacement (page 39)
step
13. Edit the text using the select button and arrow keys.
Chapter 5 – Maintenance
49