Table 19 Hmp115/T Probe Measurement Performance - Vaisala VaiNet RFL100 User Manual

Wireless
Hide thumbs Also See for VaiNet RFL100:
Table of Contents

Advertisement

RFL100 User Guide
Property
at 0 ... +40 °C (+32 °F ... +104 °F)
at −40 ... 0 °C, +40 ... +80 °C (−40 ... +32 °F, +104
... +176 °F)
Factory calibration uncertainty
Temperature sensor
1)
Includes non-linearity, hysteresis, and repeatability.
2)
Small variations possible; see also calibration certificate.
Table 19 HMP115/T probe measurement performance
Property
Relative humidity
Measurement range
Accuracy in temperature range 0 ... +40 °C (+32 ... +104 °F)
0 ... 90 %RH
90 ... 100 %RH
Accuracy in temperature range −40 ... 0 °C, +40 ... +60 °C (−40 ... +32 °F, +104 ... +140 °F)
0 ... 90 %RH
90 ... 100 %RH
Factory calibration uncertainty at +20 °C (68 °F)
0 ... 40 %RH
40 ... 75 %RH
Humidity sensor
Stability
Temperature
Measurement range
Accuracy over temperature range
at 0 ... +40 °C (+32 ... +104 °F)
at −40 ... 0 °C, +40 ... +60 °C (−40 ... +32 °F,
+104 ... +140 °F)
Factory calibration uncertainty
64
Description/Value
± 0.2 °C (0.36 °F)
± 0.4 °C (0.72 °F)
± 0.2 °C (0.36 °F)
2)
Pt1000 RTD Class F0.1 IEC 60751
Description/Value
0 ... 100 %RH
±1.5 %RH
±2.5 %RH
±3.0 %RH
±4.0 %RH
2)
±0.6 %RH
±1.0 %RH
HUMICAPâ 180R
±2 %RH over 2 years
−40 ... +60 °C (−40 °F ... +140 °F)
± 0.2 °C (0.36 °F)
± 0.4 °C (0.72 °F)
± 0.1 °C (0.18 °F)
2)
1)
M211861EN-D
1)

Advertisement

Table of Contents
loading

Table of Contents