Advertisement

5. Define the needed adjustment for the first adjustment point:
a. Insert the probe head in the reference environment for the first adjustment point.
b. Wait for the measurement to stabilize fully.
c. Click the Reference value, point 1 text box and enter the reference value of the
adjustment point. Press ENTER or click outside the text box when done.
d. The probe automatically enters the measured values for the adjustment point.
6. If you want to adjust in more than one point, repeat
points.
7. Select Activate adjustment > Yes to store the adjustment in the probe.
8. Check the message that appears at the top of the screen. If the message indicates that
the adjustment is activated successfully, your adjustment is stored in the probe.
9. Select the Calibration information tab and update the content of the Calibration date
and Calibration text fields.
10. Select Close > Yes to exit the calibration mode.
11. Disconnect the probe from the USB cable.
12. Connect the probe to the data logger.
More information
‣ Vaisala Insight software (page 36)
‣ Probe replacement (page 39)
5.5 Changing RFL100 batteries
• 2 pcs of new AA size 1.5 V batteries: alkaline (type LR6) or lithium (type FR6)
1. Remove the data logger from the mounting bracket.
2. Open the battery cover of the data logger.
3. Move the power switch to the Off position.
4. Remove the old AA size batteries from the data logger.
5. Check the battery orientation markings on the data logger and insert the new batteries in
the correct orientation.
6. Move the power switch to the On position.
7. Close the battery cover of the data logger. Push the latch down until you hear a click. If
the cover does not close easily, push the probe (or extension cable) in and try again.
8. Insert the data logger back in the mounting bracket.
Chapter 5 – Maintenance
step 5
for all desired adjustment
53