Vaisala VaiNet RFL100 User Manual page 42

Wireless
Hide thumbs Also See for VaiNet RFL100:
Table of Contents

Advertisement

RFL100 User Guide
5.3.2 Replacing a probe when one probe is connected
1. Remove the data logger from the mounting bracket.
2. Open the battery cover of the data logger.
3. Move the power switch to the Off position.
4. To replace a fixed probe (directly connected to RFL100 without a cable):
1
2
a. Grip the old probe from above the filter and hold the data logger with your other
hand. Pull the probe straight out of the data logger. Do not rotate the probe.
b. Align the orientation mark on the new probe with the line above the probe connector.
Push the probe straight in all the way, do not rotate.
40
1
Orientation mark on data logger
2
Orientation mark on probe
M211861EN-D

Advertisement

Table of Contents
loading

Table of Contents