8.12.1 Zero Calibration of Analog Output

With this menu item you can correct the zero-point of the analog output (4mA). The zero-point correction is only
possible when the analog output is in active mode.
The error message of the output monitoring is suppressed as long as the menu Calibration AO is open.
Therefore, connect the Ampere Meter (measuring range 20 mA DC) to the analog output only after having
opened the menu.
Calibration AO 1
Connect amperemeter to the analog output.

8.12.2 Calibration of the Digital Sensors

Please refer to Honeywell Disclaimer of the sensors.

8.12.3 Credit Menu

Display of current credits, loading new credits. Is required only for 4 – 20 mA inputs.
Credit Menu
Number of Credits

8.13 Setting RS-485 Address

Addressing
Assignment of the basis slave address of the device for field bus operation
Setting Address
®
PolyGard
is a registered trademark of MSR
Phone 0049(0)8531/9004-0
Fax: 0049(0)8531/9004-54
MSR-Electronic GmbH, Bgm.-Schoenbauer-Str. 13, D 94060 Pocking www:msr-electronic.de Made in Germany
®
2
Calibration AO 1
320
0
Calibration AO 1
320
323
Calibration AO 1
SAVE
Calibration AO 1
323
323
Display of number of current credits
40
If display is 0, calibration isn't possible any more.
Define the basis slave address
4
Display of the current zero offset on the left.
Adjust the zero offset on the right by changing the
offset value slowly, until the amperemeter shows 4 mA.
Save the adjusted zero offset.
Return to the display of the current zero offset.