Electrical
Power supply
Overvoltage category
Power consumption (24 V DC)
- Control Board
- per sensor (analog)
- horn / warning light
Overvoltage category
Alarm relays (3)
Transistor output (2)
Digital input (2)
Analog input (3)
Analog output signal (1)
Ambient conditions
Temperature range
Humidity range
Storage temperature
Serial interface
Field bus
Tool bus
Physical
Housing type C
Combustion
Housing colour
Dimension housing (W x H x D)
Type C
Weight
Protection class (delivery status*)
Installation
Knockouts for cable and sensor entry
Wire connection:
Digital input, analog output
Power supply, relays
Directives
Warranty
*If there are changes on the housing it has to be re-evaluated
LCD Display
LCD
Operation
Power consumption
®
PolyGard
is a registered trademark of MSR
Phone 0049(0)8531/9004-0
Fax: 0049(0)8531/9004-54
MSR-Electronic GmbH, Bgm.-Schoenbauer-Str. 13, D 94060 Pocking www:msr-electronic.de Made in Germany
®
2
24 V DC ± 20 %, reverse-polarity protected
24 V AC ± 15 %
I
Max. 60 mA (1.5 VA), without sensor
Max. 85 mA (2.1 VA)
Max. 40 mA (1.0 VA)
I
240 V AC, 5 A, potential-free, change-over contact (SPDT)
24 V DC / 0.1 A (switching to plus) (only at 24 V DC power supply)
Potential-free
4 – 20 mA overload and short-circuit proof, input resistance 200 
Proportional, overload and short-circuit proof, load  500 Ohm
4 - 20 mA = measuring range
3.0 < 4 mA = underrange
> 20 - 21.2 mA = overrange
2.0 mA = fault (not preset)
-25 °C to +50 °C (-13 °F to +122 °F)
15 - 95 % RH not-condensing
+5 °C to +40 °C (+41 °F to +104 °F)
RS 485 / 19200 Baud
2-wire / 19200 Baud
Polycarbonate
UL 94 V2
RAL 7032 (light grey)
130 x 130 x 75 mm (5.12 x 5.12 x 2.95 in.)
ca. 0.6 kg (1.32 lb.)
NEMA 4X (IP 65)
Wall mounting
Standard 6 x M20/25
Screw-type terminal min. 0.25 mm2, max. 1.3 mm2 (24 to 16 AWG)
Screw-type terminal min. 0.25 mm2, max. 2.5 mm2 (24 to 10 AWG)
EMC directives 2014/30/EU
Low voltage directive 2014/35/EU
CE
EN 61010-1:2010
1 year on device
Two lines, 16 characters each, background highlighted in two
colours
Menu driven via six pushbuttons
5 V, 60 mA, 0.3 VA