7.2 Setting / Changing of Parameters and Set Points

Open desired menu window.
Code input field opens automatically, if required.
After input of valid code, the cursor jumps onto the first position segment to be changed.
Push the cursor onto the position segment, which has to be changed.
Set the desired parameter / set point with the keys.
Input of value finished.
Save the changed value?
YES, and back to higher menu level.
NO, (previous value isn't overwritten) and back to higher menu level.

7.3 Code Levels

All inputs and changes are protected by a four-digit numeric code (= password) against unauthorized
intervention according to the regulations of all national and international standards for gas warning systems.
The menu windows of status messages and measuring values are visible without entering a code.
The access to a code level is cancelled if no button is pushed within 15 minutes or if there is no data
communication between display and basic board.
The code levels are classified in order of priority: Priority 1 has top priority.
Priority 1: (code not changeable)
Code level priority 1 is intended for the service technician of the installer to change parameters and set-points.
This password allows working on all settings. For opening the parameter menus, you must first activate the
service mode after code release.
Priority 2: (code not changeable)
This functionality isn't available.
Priority 3: (customer password is settable)
Customer password is inactive in delivery state and is activated by entering a value. Same behaviour as priority
1 password, only changing the customer password is not possible.
Normally the code is only known by the service technician who has last changed it since it can be changed
individually via priority 1.
Priority 4: (password 1234) (code not changeable)
Code level priority 4 allows the operator after activation of the operation mode "Service Mode" to read all
parameters as well as all test functions of the alarm relays, analog outputs and LCD.
- Manual test function of the alarm relays (functional test of the connected actuators),
- Manual test function of the analog outputs (functional test of the connected actuators),
- Manual test function of the LCD (functional test of the LCD display and the LEDs).
®
PolyGard
is a registered trademark of MSR
Phone 0049(0)8531/9004-0
Fax: 0049(0)8531/9004-54
MSR-Electronic GmbH, Bgm.-Schoenbauer-Str. 13, D 94060 Pocking www:msr-electronic.de Made in Germany
®
2
Change further parameters in the same menu.