Functions
8.6 Configuration
No binary input has been preassigned in the factory.
Press the fifth softkey ("...cont."), to access the next function screen, thus branching to
further binary inputs.
Note
Save changes
Make sure you save every change in the configuration of the binary inputs in the user data
memory using function 75! If you neglect to do this, there is a danger that a previous
(unwanted) configuration will be loaded by "Load user data" (function 75).
Table 8- 6
Function
Not assigned
Fault 1 to 7
Maintenance
request 1 to 7
Acknowledge
Function control
1 to 4
AUTOCAL
Measuring range
1 to 4
Zero gas
110
Control functions of the binary inputs
Control voltage needed
0 V 24 V 24 V
(min. 1 s)
N
X
N
X
N, X
N
X
N, X
N, X
N, X
Remark/effect
No effect when input is set
e.g. due to:
Signal from a gas preparation system:
Condensate overflow
Gas cooler defective
Fault message from interfering gas analyzer
Setting the input deletes all logbook entries.
However, if the cause of a fault or maintenance re-
quest has not been eliminated, the corresponding
message reappears in the logbook.
Relay must be configured for function control under
function 71 if, for example, the function is to be con-
trolled by a second device.
Note:
Effective control only possible in "Measuring" operat-
ing state!
Autocal must be configured (function 24)
When the input is set, the corresponding measuring
range is switched to.
Note:
Effective control only possible in "Measuring" operat-
ing state!
Relay must be configured for zero gas. After the input
is set, the device changes from "Measuring" to "Cali-
brating" operating state. See also "Example of binary
inputs"