Siemens FIDAMAT 6 Operating Instructions Manual page 83

Continuous gas analysis, gas analyzer for determination of total hydrocarbon content
Table of Contents

Advertisement

Single calibration ("Measuring" operating state required)
Figure 8-4
Single calibration is only possible if the
deactivated in function 23.
In the display, the same number of measuring ranges appear as you have specified with
function 41. The above function display is an example of a single calibration of four
measuring ranges.
If, for example, you want to calibrate measuring range 3, press the corresponding softkey.
The setpoint and current actual value of the selected measuring range now appear in the
display:
Figure 8-5
When the actual value has stabilized, you can initiate the calibration by pressing the fourth
softkey. The actual value is now made to agree with the setpoint.
If you make a mistake during the calibration (e.g. wrong calibration gas), reload the original
calibration by pressing the "CANCEL" softkey.
Total range calibration ("Measuring" operating state required)
A total range calibration is only possible if the
activated under function 23.
FIDAMAT 6
Operating Instructions, 01/2019, A5E00222135-04
Performing a single calibration
Calibrating the measuring range
"Total range calibration"
"Total range calibration"
Functions
8.3 Calibration
subfunction has been
subfunction is
81