Commissioning
6.1 Safety information for commissioning
Leakiness of gas path
Danger of poisoning
Leaky gas paths lead to accumulation of the sample gas in the device.
• Tighten the unions in accordance with the mounting specifications of the manufacturer
• Use only sample gas-contacting connecting parts (pipes, unions and sealing material)
Hot surfaces due to hot sample gases and high-temperature devices
Danger of burns due to surface temperatures above 70 °C (155 °F)
The inside of the oven reaches a temperature of approximately 200 °C during operation.
After being switched off, the device and its parts are still very hot and the temperature only
drops very slowly. You can call up the actual temperature of the oven and other
temperature parameters via the Diagnostic values (function 2) (Page 77).
Contact with skin may result in burns.
• Wait until the device has cooled down before starting maintenance work or
• Take appropriate protective measures during maintenance, e.g. contact prevention.
Hot surfaces resulting from hot sample gases and heated devices
Danger of burns resulting from surface temperatures above 70 °C (155 °F).
• Take appropriate protective measures, for example contact protection.
• Make sure that the maximum ambient temperature is not exceeded by implementing
54
WARNING
using a suitable open-ended wrench.
In the process, make sure that you counter properly.
that are suitable for the connection and for the sample gases.
CAUTION
decommissioning.
CAUTION
protective measures. Refer to the information in the section "Technical specifications
(Page 167)".