8.6 Configuration
Relay and binary test
Figure 8-56
The first function screen shows the state of the six relay and binary inputs of the standard
device. With an optional board, there are another eight relay/binary inputs on a second page.
With the relay test, individual relays can be activated. This is done via the input field. With
"1", the relay is on, with "0", it returns to its release condition. Numbers other than "0" and "1"
are not accepted by the input field.
After leaving function 80, the relays are back in the state they had before calling the relay
and binary test.
The current state of the binary inputs is displayed in the "Binary" column.
Analog test
With the analog test, the analog output is set to a constant current of 0 to 24000 µA for
testing purposes.
The analog input displays the input currents in µA.
8.6.13

Language selection (function 81)

Figure 8-57
122
Test relay outputs and binary inputs
Selecting a language