Language Selection (Function 81); Correction Of Cross-Interference (Function 83) - Siemens OXYMAT 64 Operating Instructions Manual

Gas analyzer for measurements of trace oxygen
Hide thumbs Also See for OXYMAT 64:
Table of Contents

Advertisement

5.4.6.13

Language selection (function 81)

81 Select language
Figure 5-61
With this function, you set the device to a second dialog language.
The device is always delivered in the ordered language. Usually, English is included as a
second language (if English is the first language, Spanish is set as the second language).
5.4.6.14

Correction of cross-interference (function 83)

83 Interf. correct.
Interference correct. 1
Interference correct. 2
Interference correct. 3
Interference correct. 4
Figure 5-62
Flammable gases influence the O
sample gas. The sensor temperature, the concentration of flammable components, as well
as the O
The relationship between interfering gas concentration and O
(see also Table "Examples of typical interfering gas offsets on a catalytically inactive
sensor").
Note
A differentiation must always be made with the correction of cross-interference as to whether
the interfering gas has a constant or variable concentration.
OXYMAT 64
Operating Instructions, 06/2008, A5E00880383-03
other
English
Selecting a language
Carrying out correction of cross-interference
concentration and sample gas flow all have an effect.
2
O
2
O
2
measurement in that they reduce the O
2
Operation
5.4 Operating functions
content of the
2
offset is usually non-linear
2
95