Repairs
(cont.)
Draining the fuel cell module system separation circuit
Fig. 79
1. Place the fuel cell module in maintenance mode.
See page 64.
A
B
2. Turn off the ON/OFF switch on the fuel cell mod-
ule. See page 58.
3. Attach the supplied short hose to fitting
connect it to a drain.
4. Open the drain valve on the underside of fitting
by turning screw
to undo it and pulling it down-
B
wards.
Note
If necessary, open the air vent valves on the two
inner shut-off valves above the modules.
Draining the system on the DHW side
Fig. 80
1. Shut off the cold water connection.
2. To vent, open a draw-off point in the domestic sys-
tem.
3. Attach a hose to connection
drain.
4. Open drain valve
.
B
Troubleshooting