Trackball Holder (If Available); Scanner - Dentsply Sirona CEREC Omnicam AC Operating Instructions Manual

Hide thumbs Also See for CEREC Omnicam AC:
Table of Contents

Advertisement

Available languages

Available languages

Dentsply Sirona
67 15 499 D3696
D3696.201.02.04.02    09.2019
6.1.3

Trackball holder (if available)

NOTE
No residues in the spherical calotte
The spherical calotte is designed without openings and may therefore
come into contact with fluids.
➢ Ensure that no cleaning agent residues or other residues remain
in the spherical calotte.
1. Remove the ball.
2. Wipe out the spherical calotte (spherical cap).
3. Wipe down the ball
4. Insert the ball.
6.1.4

Scanner

CAUTION
Risk of injury
An evidently damaged scanner must no longer be used on patients.
If the CEREC Primescan / Primescan / CEREC Omnicam / Omnicam
scanner accidentally falls off, check to make sure that the sapphire
glass is not damaged. If the CEREC Primescan / Primescan / CEREC
Omnicam / Omnicam scanner has been damaged, it must no longer
be used on patients.
The CEREC Primescan / Primescan / CEREC Omnicam / Omnicam
scanner must be recalibrated.
NOTE
Do not sterilize the CEREC Primescan / Primescan / CEREC
Omnicam / Omnicam scanner and the scanner cable!
NOTE
The mirror sleeves cannot be sterilized in the autoclave.
NOTE
For markets where the RKI* guidelines are to be observed
The mirror sleeve is classified as a "semicritical medical device A"
according to RKI guidelines and therefore does not have to be
autoclavable.
*RKI=Robert Koch Institute, Berlin (Germany).
6 Maintenance
6.1 Cleaning, disinfection and sterilization
65