Refitting The Mirror Sleeve; Mirror Sleeve Reprocessing Overview - Dentsply Sirona CEREC Omnicam AC Operating Instructions Manual

Hide thumbs Also See for CEREC Omnicam AC:
Table of Contents

Advertisement

Available languages

Available languages

6 Maintenance
6.1 Cleaning, disinfection and sterilization
Mirror sleeve for CEREC Omnicam
Pre-Cleaning process
(while mirror sleeve is attached to the camera)
1. Clean with a cotton gauze moistened with tap water of potable water quality that meets Federal Clean Water
Standards and pH neutral soap
or
use a clean lint-free cloth which has been soaked in CaviCide
or
use Cavi Wipes
or
use pdCARE Wipes
or
use a clean lint-free cloth which has been soaked in > 60% isopropyl alcohol.
2. Wipe with the absorbent cotton gauze dipped in drinking water.
3. Dry the mirror sleeve with a clean lint-free cloth.
HLD process
®
1. Use CIDEX 
OPA as disinfectant for high-level
disinfection per manufacturer´s instructions.
2. Leave the mirror sleeve in CIDEX 
minutes at 20°  C .
3. Rinse the sleeve under tap water of potable water
quality that meets Federal Clean Water Standards.
4. Dry the mirror sleeve with a clean lint-free cloth.
68
6.1.4.4

Refitting the mirror sleeve

➢ Carefully refit the mirror sleeve until it locks in place.
NOTE
There is a risk of damaging the scanner window or the coated
sapphire glass if the mirror sleeve is not pushed straight toward the
front.
➢ The mirror sleeve must not come into contact with the scanner
window.
➢ Push the mirror sleeve straight toward the scanner body; do not
tilt it.
➢ Carefully refit the mirror sleeve until it locks in place.
6.1.4.5

Mirror sleeve reprocessing overview

See detailed instructions after summary.
Dry heat sterilization process
➢ Dry heat 160°C (SteriSURE) for 60  m inutes
®
OPA for 12
(wrapped or unwrapped)
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for 6
minutes (unwrapped),
or
Dry heat 190°C (Cox RAPIDHEAT Sterilizer) for 12
minutes (wrapped).
Dentsply Sirona
67 15 499 D3696
D3696.201.02.04.02    09.2019

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cerec primescan acPrimescan acOmnicam ac

Table of Contents