6 Maintenance
6.1 Cleaning, disinfection and sterilization
70
6.1.4.8
High-level disinfection of the mirror sleeve (with HLD Set)
NOTE
Dry heat sterilization and high-level disinfection must not be
combined.
The complete process for high-level disinfection is as follows:
With HLD Set
A HLD set to support the HLD process should be ordered from Dentsply
Sirona with REF 66 05 120.
A
HLD container
B
Lid
1. Pre-cleaning (see „Pre-cleaning of the mirror sleeve").
2. Put on Personal Protective Equipment at the latest when handling
the high-level disinfectant.
3. Remove the sleeve from the Omnicam / Primescan (see „Removing
the mirror sleeve").
NOTE
It is recommended that you never switch between disinfecting and
sterilizing solutions. However if you do, make sure you thoroughly
rinse the HLD set before switching.
4. Place the white protective sleeve on the scanner tip and place the
scanner in the scanner cradle.
5. Use the following disinfectant for the high-level disinfection: CIDEX
®
OPA.
C
Metal holder
D
Plug
D3696.201.02.04.02    09.2019
Dentsply Sirona
67 15 499 D3696