Save, Recall, And The Unique User Key - Keysight HSA N9342C Demonstration Manual

Handheld spectrum analyzer
Table of Contents

Advertisement

09 | Keysight | N9342C/N9343C/N9344C HSA Handheld Spectrum Analyzer - Demonstration Guide
Demonstration 3

Save, recall, and the unique User key

The N934xC HSAs let you save the state, trace data, measurement
results (peak and marker table), limit lines, and screen captures to an
internal file, a USB drive, or remotely via LAN or USB.
The N934xC HSAs provide a dedicated Save button to let you save a
file quickly with a single button press. When the Save button is used,
you can choose to name the file by yourself, or let the HSA automati-
cally pick a file name that consists of a prefix (N934xC is the default
but can be edited) and a index.
Another unique feature on the N934xC HSAs is the User key, which
provides a quick way to set up the instrument. You can select from
7 predefined instrument setups with a single button push.
Instructions for the source
Set 2-GHz center frequency, amplitude -10 dBm, W-CDMA signal
Instructions for HSA
Preset N934xC HSA
Select internal C:/USER/STATE folder
Note: For the user key feature, the STA file must be placed under the
USER/STATE directory
You may have to create the USER and STATE directory folders
Set QSave file type
Turn on QSave naming
Set center frequency and span
Turn on channel power
Set integrated BW
Save file
Turn on OBW
Save file
Preset N934xC HSA
Use the User key
(As shown in Figure 17, channel power and OBW's state file name appears
on the soft key button)
Make measurement with User key
Channel power
OBW
Figure 3. USER key
Keystrokes for the source
[Preset], [Freq], [2], {GHz}, [Amplitude], [-10], {dBm}, [Mode], {W-CDMA}, {Arb
W-CDMA (3GPP 12-2004}, {W-CDMA (On)}, [Mod On], [RF On]
Keystrokes for HSA
[Preset]
[Shift], [File], rotate the knob to select the USER folder, [Enter], rotate the
knob to select the STATE folder, [Enter], {Return}
[Shift], [File], {Directory}, {Create Folder}
[Shift], [File], {Setup}, {QSave Type}, {STA}
{QSave Naming (On)}, {Return}
[Freq], {Center Freq}, [2], {GHz}, [Span], [10], {MHz}
[Meas], {Channel Power}
{Integrated BW}, [5], {MHz}
[Shift], [Save], input the file name [WCDMA_CHP] in the "Save As" dialog box
and press [Enter]
[Meas], [Meas], {Occupied BW}
[Shift], [Save], input the file name [WCDMA_OBW] in the "Save As" dialog box,
[Enter]
[Preset]
[Shift], [User]
{WCDMA_CHP}
[Shift], [User], {WCDMA_OBW}

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hsa n9344cHsa n9343c

Table of Contents