21 | Keysight | N9342C/N9343C/N9344C HSA Handheld Spectrum Analyzer - Demonstration Guide
Demonstration 11

Multiple traces and simultaneous detectors

The N934xC HSA has a total of four traces and supports simultaneous detectors
including peak, average, sample, negative peak, and the normal detector. Plus the
N934xC HSA supports simultaneous detectors, meaning that different detectors can be
used on four different traces in a single sweep—this will greatly increase the throughput
to measure crest factor of pseudo-noise signals, as well as amplifiers.
In this demonstration, we will turn on three traces with peak, RMS, and neg peak
detectors.
Figure 24. Multiple traces and simultaneous detectors
Instructions for the source
Set 2 GHz center frequency, amplitude -10 dBm, W-CDMA signal
Instructions for HSA
Preset, set the center frequency, span, and RBW
Set reference level
Set Y scale
Set peak detector for trace 1 and turn on average function.
Set neg peak detector for trace 2 and turn on average function
Set RMS detector for trace 3 and turn on average function
Keystrokes for the source
[Preset], [Freq], [2], {GHz}, [Amplitude], [-10], {dBm}, [Mode], {W-CDMA},
{Arb W-CDMA (3GPP 12-2004}, {W-CDMA (On)},
[Mod On], [RF On]
Keystrokes for HSA
[Preset], [Freq], {Center Freq}, [2], {GHz}, [Span], [10], {MHz}, [BW], {RBW}
(Man), {10 kHz}
[Amptd], [-30], {dBm}
[Amptd], {Scale/DIV}, {2/DIV}
[Trace], {More 1 of 2}, {Detector}, {Pos Peak}, {Return}, {Average Number
(On)}, {More 2 of 2}
{Trace (2)}, {Clear Write}, {More 1 of 2}, {Detector}, {Neg Peak}, {Average
Number 100 (On)}
[BW], {Average Type}, {Power}, {Trace (3)}, {Clear Write}, {More 1 of 2}, {Detec-
tor}, {Average (RMS)}, {Average Number (On)}