16 | Keysight | N9342C/N9343C/N9344C HSA Handheld Spectrum Analyzer - Demonstration Guide
Demonstration 5
Task planner
(continued)
Figure 17. Channel power measurement result
Figure 18. OBW measurement result
Step 3: Run task planner
Instructions for the source
Set 2 GHz center frequency, amplitude -10 dBm,
W-CDMA signal
Instructions for HSA
Load the task plan file
Run the task plan file (Figure 15)
Run task 1: Channel power
Run task 2: OBW
Run task 3: ACPR
Figure 19. ACPR measurement result
Keystrokes for the source
[Preset], [Freq], [2], {GHz}, [Amplitude], [-10], {dBm}, [Mode], {W-CDMA}, {Arb W-CDMA (3GPP 12-
2004}, {W-CDMA (On)}, [Mod On], [RF On]
Keystrokes for HSA
[Shift], [File], use the rotating knob to choose W-CDMA.tpf, [Enter]
{Run}
– Press [Enter] after the "chp begin Enter/Esc to Confirm" message appears
– Press [Enter] after the "chp end Enter/Esc to Confirm" message appears
– Press [Enter] after the "obw begin Enter/Esc to Confirm" message appears
– Press [Enter] after the "obw end Enter/Esc to Confirm" message appears
– Press [Enter] after "acpr begin Enter/Esc to Confirm" message appears
– Press [Enter] after "acpr end Enter/Esc to Confirm" message appears
N934xC HSA automatically creates a folder whose name is the same as the task plan file name (for
this demo, the directory is named as W_CDMA) to save all report contents, including a text report,
screenshots, and trace files depending on the choices of report contents. The PASS/FAIL result is
shown in the report (Figures 17 to 19)