7.18 Setting the Console Speed Step Resolution

Extended Menu
Pressure Display
Speed Step Resolution
Flow Range
Flow Limit Sensitivity
Flow Recorder Speed*
Support Type
Language
Data Logger*
Copy Data*
Management*
* Monitor only
The System may be used to treat patients with a wide range of body sizes and conditions.
The SPEED STEP RESOLUTION function is provided to allow the user to select smaller
incremental changes in speed (50 RPM increments) for small individuals, and larger (100
RPM) changes for the larger patients. The default setting is the 100 RPM option. To
access the SPEED STEP RESOLUTION function, depress the MENU keypad and scroll
down to the EXTENDED MENU option, then depress the Menu key until the SPEED
STEP RESOLUTION option appears. Choose between the STEP 50 (50 RPM) and STEP
100 (100 RPM) options using the UP and DOWN arrows. Once selected, this will dictate
the speed that the Pump will increase with each depression of the UP or DOWN ARROW
keypads when using the SET RPM option.
TM
CentriMag
Circulatory Support System Operation Manual (US)
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Figure 46: Monitor MENU Structure
Speed Step Resolution
Set RPM
Set RPM
Exit
Set RPM
Inactive
Fine
Set RPM
(50 rpm)
Reduced
Set RPM
(0 – 3 lpm)
Set RPM
Sensitive
Slow
Set RPM
(2 min)
Set RPM
L
Set RPM
Select
Set RPM
Set RPM
[Execute|Cancel]
Set RPM
[Start|Cancel]
Return
Active
Normal
(100 rpm)
Normal
(0 – 10 lpm)
Normal
Fast
(20 sec)
R
Next
Enter
[
|Confirm]
[
|Confirm]
Page 75 of 126