Specifications; Environmental Conditions - Abbott CentriMag Operation Manual

Circulatory support system
Hide thumbs Also See for CentriMag:
Table of Contents

Advertisement

6.2 Specifications

PARAMETER
AC Power
Battery Power
Dimensions
Weight
Pump Speed Range
Pump Flow Range
Flow Range Display
Electrical Safety

6.3 Environmental Conditions

6.3.1 Shipping Conditions
The following are the acceptable environmental conditions during shipping:
Temperature: -29
Relative humidity: 0% to 85%
Atmospheric pressure: 210 hPa – 1100 hPa (157 mmHg – 825 mmHg)
6.3.2 Operational and Storage Conditions
The following are the acceptable environmental conditions during operation and storage:
Temperature: 10
Relative humidity: 30% to 75%
Atmospheric pressure: 702 hPa – 1100 hPa (527 mmHg – 825 mmHg)
1
Note: If the probe detects forward flow of more than 10 LPM then it will display as "^^.^^ LPM".
TM
CentriMag
Circulatory Support System Operation Manual (US)
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
Table 5: Console Specifications
100 – 240 VAC
at 50/60 Hz, 170 VA
14.8 VDC Li-Ion, internal rechargeable battery
Discharge time: approx. 120 minutes @3,500 RPM,
Recharge time: 4 hrs to 90% charge, 5 hrs to 100%
Height: 10.0 cm / 3.9 in
Width: 26.6 cm / 10.5 in
Depth: 33.0 cm / 13.0 in
5.8 kg / 12.8 lbs
0 – 5,500 revolutions per minute (RPM)
0.0 – 10.0 liters per minute (LPM)
-2.0 – 10.0 liters per minute (LPM)
Earth leakage current: < 500 µA
Touch current: < 100 µA
Patient leakage current: < 10 µA
o
o
C to 60
C (one week maximum)
o
o
C to 30
C
SPECIFICATIONS
5.5 LPM
charge
1
Page 34 of 126

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents