12.2 Appendix II – Technical Specification
nd
2
GENERATION CENTRIMAG CONSOLE
PARAMETER
AC Power
Battery Type & Chemistry
Battery Voltage
Available Battery Run
Time
Battery Recharge Time
Dimensions
Weight
Pump Speed Range
Pump Flow Range
Flow Range Display
Flow Resolution
Pressure Range Display
Alarm/Alert Audio Volume
Language Options
Electrical Safety
Fuse
Shipping Condition
Operational & Storage
Conditions
nd
2
Generation CentriMag
Console Application
Software
TM
CentriMag
Circulatory Support System Operation Manual (US)
© 2019 Abbott– Document No PL-0047, Rev 11 (December 2019)
100 – 240 VAC
at 50/60 Hz, 170 VA
Rechargeable internal battery, Lithium Ion
14.8 Volts
approx. 120 min @ 3,500 RPM, 5.5 LPM
4 hrs to 90% charge, 5 hrs to 100% charge
Height: 10.0 cm / 3.9 in
Width: 26.6 cm / 10.5 in
Depth: 33.0 cm / 13.0 in
5.8 kg / 12.8 lbs
0 – 5,500 revolutions per minute (RPM)
0.0 – 10.0 liters per minute (LPM)
-2.0 – 10.0 liters per minute (LPM)
10 mLPM
-150 – 900 mmHg
70 dB
English, Dutch, French, German, Italian and Spanish
Earth leakage current: < 500 µA
Touch current: < 100 µA
Patient leakage current: < 10 µA
"5 x 20 mm, T 3.15A H 250V"
o
Temperature: -29
C to 60
Relative humidity: 0% to 85%
Atmospheric pressure: 210 hPa – 1100 hPa (157 mmHg –
825 mmHg)
o
Temperature: 10
C to 30
Relative humidity: 30% to 75%
Atmospheric pressure: 702 hPa – 1100 hPa (527 mmHg –
825 mmHg)
Version CPC1.02
LMCEBPX:
IFD:
SPS-MSP:
SPS-PIC:
SPECIFICATIONS
o
o
C / -20
F to 140
o
o
o
C / 50
F to 86
F
SW-0032-01
SW-0033-01
SW-0034-01
SW-0043-01
o
F
Rev 00
Rev 02
Rev 02
Rev 00
Page 118 of 126