7.16 Flow range

Extended Menu
Pressure Display
Speed Step Resolution
Flow Range
Flow Limit Sensitivity
Flow Recorder Speed*
Support Type
Language
Data Logger*
Copy Data*
Management*
* Monitor only
The System may be used to treat a variety of conditions to accommodate different ranges
of flow. Under different conditions, the System may be operated in one of two flow ranges
0 – 10 LPM and 0 – 3 LPM. Changing the flow range will adjust the range of the flow
display bar, as well as the default limit for the maximum flow alarm if the high limit is more
than 3.0 LPM. The most common use of the low flow range is for display of flow during
use of the smaller flow transducer designed to be placed on ¼" ID tubing.
To change between the two ranges, depress the MENU button until the EXTENDED
MENU option is displayed, then use the MENU key to scroll through the menu options until
the FLOW RANGE option is displayed. Select between the two options using the arrow
keys.
TM
CentriMag
Circulatory Support System Operation Manual (US)
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Menu
Figure 44: Monitor MENU Structure
Flow Range
Set RPM
Set RPM
Exit
Set RPM
Inactive
Fine
Set RPM
(50 rpm)
Reduced
Set RPM
(0 – 3 lpm)
Set RPM
Sensitive
Slow
Set RPM
(2 min)
Set RPM
L
Set RPM
Select
Set RPM
Set RPM
[Execute|Cancel]
Set RPM
[Start|Cancel]
Return
Active
Normal
(100 rpm)
Normal
(0 – 10 lpm)
Normal
Fast
(20 sec)
R
Next
Enter
[
|Confirm]
[
|Confirm]
Page 73 of 126