4 WARNINGS AND PRECAUTIONS

WARNING
Read this entire manual before you use the CentriMag System. As with all
prescription medical devices, clinical procedures should be conducted under
the direction of the prescribing physician. The physician must be trained on the
use of the System before using it. The professional staff at Abbott Medical
regularly provide laboratory training and on-site, in-service programs.
information,
please
representative.
WARNING
The System is designed to be operated only with the CentriMag and PediMag
Pumps. There are no safety or performance data that establish compatibility
with any other manufacturer's device or components.
WARNING
One additional CentriMag Console, Motor and Flow Probe are required as
backup components in the immediate vicinity of each patient whenever the
CentriMag or PediMag Pump is used. The backup Console must be connected to
the backup Motor and to the backup Flow Probe, have a battery charge
sufficient for at least one hour of operation, be connected to AC power (except
during transport) and be immediately available should the main Console, Motor
or Flow Probe experience a malfunction.
WARNING
The safety and effectiveness of use of the System in an ECMO circuit (i.e.
cardiopulmonary support > 6 hours) has not been demonstrated.
WARNING
The safety and effectiveness of use of the System for use > 30 days has not
been demonstrated.
TM
CentriMag
Circulatory Support System Operation Manual (US)
© 2019 Abbott – Document No PL-0047, Rev 11 (December 2019)
contact
your
local
Abbott
Medical
For
Clinical
Field
Page 11 of 126