Technical Data

15.7 Metrological Data

15.7.1 Models MCA2.7S | MCA10.6S | MCA6.6S | MCA3.6P
Scale interval (d)
Maximum capacity (Max)
Repeatability at 5 % load
Standard deviation of the load values,
tolerance
Standard deviation of the load values,
typical value
Repeatability at approx . maximum capacity
Standard deviation of the load values,
tolerance
Standard deviation of the load values,
typical value
Linearity deviation
Tolerance
Typical value
Deviation when load is off-center,
positions according to OIML R76
Test weight
Tolerance
Typical value
Sensitivity drift between +10 °C and +30 °C
Tare maximum capacity: Less than 100 % of
maximum capacity
Accuracy class according to Directive 2014/31/EU
Verification scale interval (e) according to
Directive 2014/31/EU
Minimum load (Min) according to Directive
2014/31/EU
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap . 41
Optimum minimum initial weighing
Typical minimum initial weighing
Typical stabilization time
Typical measurement time

15.8 Recommended Calibration Weight

External test weight
Recommended accuracy class
74
Cubis
®
MCA Operating Instructions
MCA2.7S
Unit
Value
µg
0 .1
g
2 .1
µg
0 .20
µg
0 .15
µg
0 .25
µg
0 .18
µg
0 .9
µg
0 .7
g
1
µg
0 .7
µg
0 .5
ppm/K
1
I
mg
1
mg
0 .01
µg
82
µg
300
s
7
s
10
MCA2.7S
Unit
Value
g
2
E2
MCA10.6S
MCA6.6S
Value
Value
1
1
10 .1
6 .1
1
1
0 .5
0 .5
1
1
0 .6
0 .6
4
4
3
3
5
2
4
4
3
3
1
1
I
I
1
1
0 .1
0 .1
820
820
1000
1000
5
5
8
8
MCA10.6S
MCA6.6S
Value
Value
10
5
E2
E2
MCA3.6P
Value
1 | 2 | 5
1 .1 | 2 .1 | 3 .1
3
0 .5
5
0 .6
4
3
1
5
3
1
I
1
0 .1
820
1000
5
8
MCA3.6P
Value
3
E2