Parameters in the "SBI Protocol" Submenu
Parameters
Configuration values
Format
Readout without identifier
Readout with identifier*
Date/time and readout
Readout with weighing
block (N, T, GC)
Output
without stability*
with stability
Automatic data
Off, triggered by means of
output
SBI commands (ESC P)*
Each weight value
(according to display result)
1 second
2 seconds
5 seconds
10 seconds
30 seconds
60 seconds
* Factory setting
Parameters in the "PC Direct Protocol" Submenu
Parameters
Configuration values
Output (TAB-
Text (identifier, value,
separated),
and unit)
multiple selection
Title (in regional language)
Identifier*
Readout*
Unit*
Decimal marker
Decimal point*
of readout
Decimal comma
* Factory setting
Explanation
The data output only exports the measured value without ID .
The data output exports the measured value with ID codes .
The data output exports the measured value, date, and time .
The data output exports the measured value with the net, tare, and
calculated gross value .
The data is output without the weight value being stable .
The data is output only when the weight value is stable .
Deactivates the automatic data output . The data output must be triggered
by means of an SBI command .
Starts the automatic data output with each value .
Starts the automatic data output after every second .
Starts the automatic data output every 2 seconds .
Starts the automatic data output every 5 seconds .
Starts the automatic data output every 10 seconds .
Starts the automatic data output every 30 seconds .
Starts the automatic data output every 60 seconds .
Explanation
The data output exports the measured value with identifier and unit
directly into a spreadsheet program .
The data output exports the title in the regional language directly into
a spreadsheet program .
The data output exports the identification codes (IDs) directly into
a spreadsheet program .
The data output exports the readout directly into a spreadsheet program .
The data output exports the displayed unit directly into a spreadsheet
program
Sets a point as a decimal separator for the data output .
Sets a comma as a decimal separator for the data output .
System Settings