Sartorius Cubis MCA Series Operating Instructions Manual page 43

Micro balances
Hide thumbs Also See for Cubis MCA Series:
Table of Contents

Advertisement

Parameters
Configuration values
Application filter
Weighing (final readout)*
Dosing (initial weighing)
Sensor mode (low filtering)
Dynamic mode
(without filtering)
Stability
Very high degree of
accuracy
High degree of accuracy
Average degree of accuracy* Sets the stability to "average degree of accuracy" .
Fast
Very fast
Very slow
Stability delay
Very short
Short*
Medium
Long
Zeroing/taring
Without stability
With stability*
At stability
Automatic zeroing
On*
Off
Tare preset 1
On*
Off
* Factory setting
Explanation
Activates a filter that enables a fast change in the display for very fast load
changes .
Display changes with minimal load changes (in the digit range) occur more
slowly .
Activates a filter that enables a very fast change in the display with
minimal load changes (e .g . when filling containers) .
Activates a weak but fast filter that always behaves in the same way for
load changes (e .g . when filling automated systems) .
Deactivates the active application filter .
Sets the stability to "very high degree of accuracy" .
Sets the stability to "high degree of accuracy" .
Sets the stability to "fast" .
Sets the stability to "very fast" .
Sets the stability to "very slow" .
Sets the stability delay to "very short": The stability symbol is displayed
after the stability criterion is reached .
Sets the stability delay to "short": The stability symbol only appears after
a short delay in order to provide a reliable result despite fluctuations .
Sets the stability delay to "medium": The stability symbol only appears
after a longer delay in order to provide a reliable result in case of higher
fluctuations .
Sets the stability delay to "long": The stability symbol only appears after
a long delay in order to balance out major instability .
Without stability: The function of the [Zero] or [Tare] key is executed
immediately once the key is pressed .
With stability: The function of the [Zero] or [Tare] key is only executed
after stability is achieved .
At stability: The function of the [Zero] or [Tare] key is executed if stability
exists when the key is pressed .
Activates automatic zeroing . The display is automatically set to zero in case
of a deviation of 0 less than (X) .
Deactivates automatic zeroing . Zeroing must be triggered with the
[Zero] key .
Activates the initial taring / zeroing . The device is tared or zeroed after it is
switched on .
Deactivates the initial taring / zeroing . After it is switched on, the device
shows the value before it was last switched off .
System Settings
Cubis
®
MCA Operating Instructions
43