Metrological Data; Models Mca32202P | Mca70201S; Mca50201S | Mca36201S - Sartorius Cubis MCA Series Operating Instructions Manual

High-capacity precision balances
Hide thumbs Also See for Cubis MCA Series:
Table of Contents

Advertisement

Technical Data

15.7 Metrological Data

15.7.1
Models MCA32202P | MCA70201S | MCA50201S | MCA36201S
Scale interval (d)
Maximum capacity (Max)
Repeatability at 5% load
Standard deviation of the load values,
tolerance
Standard deviation of the load values,
typical value
Repeatability at approx. maximum
capacity
Standard deviation of the load values,
tolerance
Standard deviation of the load values,
typical value
Linearity deviation
Tolerance
Typical value
Deviation when load is off-center,
positions according to OIML R76
Test weight
Tolerance
Typical value
Sensitivity drift between +10°C and +30°C
Tare maximum capacity:
Less than 100% of maximum capacity
Accuracy class according
to Directive 2014 / 31 / EU
Verification scale interval (e) according
to Directive 2014 / 31 / EU
Minimum load (Min) according
to Directive 2014 / 31 / EU
Minimum initial weighing according to USP
(United States Pharmacopeia), Chap. 41
Optimum minimum initial weighing
Typical minimum initial weighing
Typical stabilization time
Typical measurement time
92
Operating Instructions Cubis
MCA32202P MCA70201S MCA50201S MCA36201S
Unit
Value
g
0.01 | 0.1
g
4200 | 32200 70200
g
0.04
g
0.02
g
0.1
g
0.02
g
0.05 | 0.2
g
0.02 | 0.1
g
10000
g
0.2
g
0.1
ppm / K
2
g
g
g
8.2
g
8.2
s
2
s
2
MCA
®
Value
Value
0.1
0.1
50200
0.1
0.1
0.04
0.04
0.1
0.1
0.04
0.04
0.5
0.5
0.15
0.15
20000
20000
0.5
0.5
0.3
0.3
2
2
II
II
1
1
5
5
82
82
82
82
1.5
1.5
1.5
1.5
Value
0.1
36200
0.1
0.02
0.1
0.02
0.2
0.1
10000
0.3
0.2
2
II
1
5
82
82
1.5
2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents