15.7 Metrological Data
15.7.1 Models MCA2.7S | MCA10.6S | MCA6.6S | MCA3.6P
Scale interval (d)
Maximum capacity (Max)
Repeatability at 5% load
Standard deviation of the load values, tolerance
Standard deviation of the load values,
typical value
Repeatability at approx . maximum capacity
Standard deviation of the load values, tolerance
Standard deviation of the load values,
typical value
Linearity deviation
Tolerance
Typical value
Deviation when load is off-center,
positions according to OIML R76
Test weight
Tolerance
Typical value
Sensitivity drift between +10°C and +30°C
Tare maximum capacity: Less than 100% of
maximum capacity
Accuracy class according to Directive 2014/31/EU
Verification scale interval (e) according to Directive
2014/31/EU
Minimum load (Min) according to Directive
2014/31/EU
Minimum initial weighing according to USP (United
States Pharmacopeia), Chap . 41
Optimum minimum initial weighing
Typical minimum initial weighing
Typical stabilization time
Typical measurement time
15.8 Recommended Calibration Weight
External test weight
Recommended accuracy class
MCA2.7S
MCA10.6S
Unit
Value
Value
µg
0 .1
1
g
2 .1
10 .1
µg
0 .20
1
µg
0 .15
0 .4
µg
0 .25
1
µg
0 .18
0 .6
µg
0 .9
4
µg
0 .7
3
g
1
5
µg
0 .7
4
µg
0 .5
3
ppm/K
1
1
I
I
mg
1
1
mg
0 .01
0 .1
µg
82
820
µg
300
820
s
7
5
s
10
8
MCA2.7S
MCA10.6S
Unit
Value
Value
g
2
10
E2
E2
Technical Data
MCA6.6S
MCA3.6P
Value
Value
1
1 | 2 | 5
6 .1
1 .1 | 2 .1 | 3 .1
1
3
0 .4
0 .4
1
5
0 .6
0 .6
4
4
3
3
2
1
4
5
3
3
1
1
I
I
1
1
0 .1
0 .1
820
820
820
820
5
5
8
8
MCA6.6S
MCA3.6P
Value
Value
5
3
E2
E2
Cubis
®
MCA Operating Instructions
83