Product Description; Functions - Dentsply Sirona DAC Universal Touch Operating Instructions Manual

Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions DAC Universal
6681006 D3708
D3708.201.01.01.02    04.2019
3

Product description

Functions

3.1
WARNING
Risk of contamination!
Only the loads listed may be reprocessed in the DAC Universal.
The DAC Universal cleans, lubricates and disinfects up to 6 turbines
and straight and contra-angle handpieces automatically in approx. 15
minutes. The dental handpieces are then immediately ready for use.
IMPORTANT
It is not possible to reprocess 6 instruments without spray channels
simultaneously on the Blue lid, as this will result in error 1506.
The number of different adapters is optional.
Please refer to the adapter operating instructions for details on the
placement of the different adapters.
For the cleaning and disinfection of up to 6 unwrapped instruments
(Green lid instruments), the Green lid is optionally available.
The following instruments can be reprocessed on the Green lid:
● Dental sonic handpieces and air scalers
● Dental ultrasonic handpieces
● Tips for dental sonic handpieces and air scalers including the
associated (torque) wrenches
● Dental ultrasonic tips including the associated (torque) wrenches
● Nozzles for multifunctional syringes
● Nozzles for powder jet devices and powder jet handpieces
The number and placement of the different adapters is optional.
Exception: A maximum of 3 ultrasonic tips from Sirona, EMS and/or
Satelec can be positioned simultaneously on the Green lid. The
ultrasonic tip adapters cannot be positioned directly next to each other.
Other Green lid adapters must be installed between the ultrasonic tip
adapters.
3 Product description
3.1 Functions
19

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents