5 Cleaning, maintenance and validation

5.3 Install a new lid

SN
8888
106
Install a new lid
5.3
If the lid is closed, the DAC Universal begins to recognize the lid.
The display shows the lid type and the serial number.
04:00 PM
To confirm a new lid press the function key " Confirm ".
IMPORTANT
Make sure that the reprocessing processes are also validated with
your new lid type if this type of lid has not yet been validated.
Observe any possible additional national requirements.

Mounting adapters

5.4
Up to 6 instruments can be reprocessed simultaneously in the DAC
Universal. The instruments are connected to the lid using adapters.
These adapters are not yet mounted when the unit is delivered.
An overview of all adapters, a detailed procedure for installing the
adapters on the respective lids and a detailed procedure for installing
the instruments on the adapters can be found in the separate operating
instructions for adapters.

Adding/replacing the lubrication concentrate

5.5
General information about the NitramOil #2 lubrication concentrate
bottle can be found in the chapter "NitramOil #2 lubrication concentrate 
[→ 26]".
Insert the oil bottle
ü The lid on the DAC Universal is open.
ü The red transport plug on the oil dispenser has been removed.
1. Remove the lid from the new NitramOil #2 bottle.
Dentsply Sirona
Operating Instructions DAC Universal
6681006 D3708
D3708.201.01.01.02    04.2019