Connecting The Suction Device (Optional) - Dentsply Sirona CEREC Primemill Operating Instructions Manual

Hide thumbs Also See for CEREC Primemill:
Table of Contents

Advertisement

5 Installation and startup
5.4 Commissioning
28
5.4.6.4

Connecting the suction device (optional)

5.4.6.4.1
Connecting the suction tube
A
CAUTION
Trip/fall hazard
If the suction tube is routed poorly, there may be a risk of tripping.
➢ To prevent injuries caused by tripping, route the suction tube so
that there is no risk of tripping.
1. Connect one end of the suction tube to the available connection
point on the suction device (A).
2. Connect the other end of the tube to the rear side of the production
unit (B).
Notes on the suction tube:
The suction tube is supplied at a length of approx. 2.0 m. When
connecting the suction device to the unit, please ensure that no sharp
bends occur over the full length of the suction tube.
Reduce the length of the tube according to your requirements and your
installation location. Note that suction power drops along the length of
the tube. You obtain decent suction power if the suction device is
placed immediately below the machine and you have a tube length of
1.2 m or less.
Dentsply Sirona
Operating Instructions CEREC Primemill
B
67 19 681 D3692
D3692.201.01.01.02    01.2020

Hide quick links:

Advertisement

Table of Contents
loading