Cleaning Surfaces; Replacing The Main Fuses - Dentsply Sirona CEREC Primemill Operating Instructions Manual

Hide thumbs Also See for CEREC Primemill:
Table of Contents

Advertisement

Dentsply Sirona
Operating Instructions CEREC Primemill
67 19 681 D3692
D3692.201.01.01.02    01.2020

Cleaning surfaces

7.5
NOTE
Do not allow liquids to run into the ventilation slots!
Remove dirt regularly using mild, commercially available cleaning
agents.
Cleaning agents categorized as "nonionic and anionic surfactant-based
soap solutions" can be used.

Replacing the main fuses

7.6
WARNING
Electric shock
Disconnect the power plug at the unit end before replacing the fuses.
NOTE
Fuse type
Use only fuses of the same type in the fuse holder!
Fuse holder
A
Cover
B
Main switch
Fuses:
ü The power plug must be disconnected.
1. Use a screwdriver to carefully pry off the cover of the fuses on the
back side of the unit.
2. Pull out the fuse holder.
3. Replace the defective fuses.
4. Reinsert the fuse holder.
5. Close the cover.
C
Fuse holder
D
Fuse
T3.15 250V
Order no. 64 45 378
7 Service
7.5 Cleaning surfaces
55