Installation

Integrating the Detector into the LC System

Integrating the Detector into the LC System
14
Surveyor RI Plus Detector Hardware Manual
Place the detector on a bench top as close as possible to the chromatographic column outlet
(minimizing the length of tubing necessary for connection to the flowcell inlet). If you are
using other Surveyor Plus modules, place the Surveyor RI Plus Detector on top of the
Surveyor Plus Autosampler as shown in
Be sure to place the HPLC system in a draft-free location away from an open window, air
conditioner vents, or other circulating air source. A stable room temperature (± 0.5 °C) is
necessary for applications requiring maximum detection sensitivity. Allow at least 15 cm
(6 in.) of clear space between the back panel of the detector and any wall or obstruction. This
provides access to the back-panel connectors and a free flow of cooling air.
Figure 9.
Surveyor Plus LC system
Power switches
Figure
9.
Detector
Power
Comm
Run
Temp
LEDs
Autosampler
Power
Comm
Run
Temp
LEDs
Pump
Power
Comm
Run
Degas
LEDs