INSTALLING THE EXPOSURE SWITCH
7.2

Stationary exposure switch

Use the holder plate to mark the position of the exposure switch.
If the wall is made of concrete or brick, use the 4x30 screws and the 6x30 expansion
anchors. Drill ø6mm (0.23 in.), 20 mm (0.8 in.) in depth, securing holes and place the
expansion anchors onto the holes.
If the wall is made of wood, use the 4x30 screws. Do not use the expansion anchors with
wooden wall. Drill ø3 mm (0.11 in.), 20 mm (0.8 in.) in depth, holes for the attachment
screws.
Make sure that the exposure switch attachment nut is in position at the lower end of the
holder plate. Attach the holder plate to the wall with the two attachment screws.
Attach the fastening plate to the exposure switch with an EJOT PT 30x8 WN1451 screw.
The exposure switch can now be positioned on the holder plate. Then attach the exposure
switch to the holder plate with an M3x10 DIN 912 screw.
B
Fastening plate
20
Planmeca ProOne
Printed copies of this document are considered uncontrolled.
7535.13.Rev001
08.07.2018
A
Holder plate
C
Exposure switch attachment nut
D