Attaching The X-Ray Unit To The Floor - Planmeca ProOne Installation Manual

Hide thumbs Also See for ProOne:
Table of Contents

Advertisement

NOTE Use a spirit level to ensure that the column is vertical.
!
6.2

Attaching the X-ray unit to the floor

NOTE The X-ray unit must be attached to the wall with at least one wall bracket. If only one
Installation Manual
Printed copies of this document are considered uncontrolled.
7535.13.Rev001
08.07.2018
Carefully move the X-ray unit against the wall bracket. Attach the wall bracket to the X-ray
unit column by tightening the four attachment screws (two on each side). The wall bracket
clamps must press firmly against the guide rails inside the column.
Attachment screws
(two on each side)
(view as seen from the side of the wall)
wall bracket is used, the X-ray unit must be attached to the floor as well. If the X-ray
unit is not attached to the floor, a second wall bracket must be used. For location of
the second wall bracket, see installation drawing in section 10 "DIAGRAMS" on page
27.
Mark the position of the two mounting screws on the floor. Then loosen the four wall
bracket attachment screws and carefully pull the X-ray unit away from the wall so that you
can drill the holes.
INSTALLING THE X-RAY UNIT
Clamp
Guide rail
Mark the position
of the screws
Planmeca ProOne 15