Installing The Digital System; Connecting The Ethernet Cable - Planmeca ProOne Installation Manual

Hide thumbs Also See for ProOne:
Table of Contents

Advertisement

INSTALLING THE DIGITAL SYSTEM

8
INSTALLING THE DIGITAL SYSTEM
NOTE If a Planmeca Dixi Ethernet system needs to be connected to the same PC together
with the ProOne/Ethernet configuration then a Hub/Switch is needed. Refer to
Planmeca Dixi User's and Installation manual.
8.1

Connecting the Ethernet cable

The ProOne X-ray unit is connected to the local (10/100 Base) network and used from a
PC with the Dimaxis/Romexis imaging software. Use an Ethernet cable (UTP cat5e,
Planmeca part number 06320060) to connect the ProOne X-ray unit to the network.
Connect the Ethernet cable to the terminal on the back of the moving column. To do this,
remove the cover that protects the terminals if it is not already off. Remove the protective
cover by unscrewing the two screws at the bottom and then unlatching the cover at the
top.
First attach a ferrite bead (included in the delivery) to the Ethernet cable. Then connect the
cable to the right hand side terminal (marked ETH).
NOTE Only the Ethernet cable can be connected to the right hand side terminal. Do not
connect any other cables to this terminal.
NOTE The left hand side terminal is a USB port and it is reserved for a USB memory stick
(future use). Do not plug any other USB devices into the USB port.
Back of
moving column
NOTE Reattach the cover that protects the terminals by fastening the two screws at the
bottom and then latching the cover in place at the top.
22
Planmeca ProOne
Printed copies of this document are considered uncontrolled.
7535.13.Rev001
08.07.2018
Screws that hold the protective cover in position
Ferrite bead
Ethernet cable
(UTP cat5e)
Installation Manual