NOTE Only the exposure switch cable can be connected to the terminals in the middle. Do
NOTE The left hand side terminal is a USB port and it is reserved for a USB memory stick
C
Back of
moving column
Installation Manual
Printed copies of this document are considered uncontrolled.
7535.13.Rev001
08.07.2018
Connect the exposure switch cable to the terminal on the back of the moving column. To
do this, you will first have to remove the cover that protects the terminals. Remove the
protective cover by unscrewing the two screws at the bottom and then unlatching the
cover at the top.
First attach a ferrite bead (included in the delivery) to the exposure switch cable. Then
connect the exposure switch cable to either of the two terminals in the middle (marked
EXP).
not connect any other cables to these terminals.
(future use). Do not plug any other USB devices into the USB port.
Screws that hold the protective cover in position
Connect the other cable end to the exposure switch.
D
INSTALLING THE EXPOSURE SWITCH
Planmeca ProOne 19
Ferrite bead
Exposure
switch cable