Keysight E8362-60109 Installation Note page 9

Extended frequency range upgrade kit

Advertisement

Step 2. Remove the Outer Cover
This procedure is best performed with the analyzer resting on its front
CAUTION
handles in the vertical position. Do not place the analyzer on its front panel
without the handles. This will damage the front panel assemblies.
1. Disconnect the power cord (if it has not already been disconnected).
2. Using a T-20 TORX driver, remove the strap handles (item
y
(item
) on both ends until the handle is free of the analyzer.
3. Using a T-20 TORX driver, remove the four rear panel feet (item
{
screws (item
).
4. Slide the four bottom feet (item
5. Slide the cover off of the frame.
Figure 2
Outer Cover Removal
Installation Note E8362-90004
|
) off the cover.
x
) by loosening the screws
z
) by removing the center
9