Download Print this page

Keysight E8362-60109 Installation Note page 7

Extended frequency range upgrade kit

Advertisement

Overview of the Installation Procedure
Step 1. Verify the Model, Serial, and Option Numbers of the Analyzer
Step 2. Remove the Outer Cover
Step 3. Remove the Inner Cover
Step 4. Remove the Front Panel Assembly
Step 5. Disconnect the A12 Source 20 to A22 Switch/Splitter Cable at the A12 Source 20
Step 6. Raise the Receiver Deck
Step 7. Remove the Old Hardware and Cables
Step 8. Replace the A38 and A39 Bias Tees (Option UNL Only)
Step 9. Install the Remaining New Hardware
Step 10. Install the Remaining Cables
Step 11. Lower the Receiver Deck
Step 12. Connect the A12 Source 20 to A21 SOMA 50 Cable at the A12 Source 20
Step 13. Replace the Front Panel Nameplate
Step 14. Reinstall the Front Panel Assembly
Step 15. Reinstall the Inner and Outer Covers
Step 16. Enter the New Model Number
Step 17. Verify that the New Model Number Is Shown Correctly
Step 18. Perform Source and Receiver Calibration Adjustments
Tools Required for the Installation
T-8 TORX driver (set to 5 in-lbs)
T-10 TORX driver (set to 9 in-lbs)
T-20 TORX driver (set to 21 in-lbs)
5/16-inch torque wrench (set to 10 in-lbs)
1-inch torque wrench (set to 72 in-lbs)
ESD Equipment and Supplies Required for the Installation
ESD grounding wrist strap
5-ft grounding cord for wrist strap
2 x 4 ft conductive table mat and 15-ft grounding wire
ESD heel strap (for use with conductive floors)
Installation Note E8362-90004
Description
Description
Agilent Part Number
N/A
N/A
N/A
N/A
N/A
Agilent Part Number
9300-1367
9300-0980
9300-0797
9300-1308
7