Keysight E8362-60109 Installation Note page 23

Extended frequency range upgrade kit

Advertisement

Step 12. Connect the A12 Source 20 to A21 SOMA 50 Cable at the A12
Source 20
1. Place the analyzer top-side up on a flat surface.
2. Using a 5/16-inch wrench, connect the A12 source 20 to A21 SOMA 50 cable, W1, at the
A12 source 20 as shown.
Figure 16 A12 Source Cable Connection
Installation Note E8362-90004
23